ELG

Medical transformers

ELG
ce

Zastosowanie

The ELG - Earth Leakage Guard - controls the dielectric resistance between the medical supply circuit and the earth potential.
The typically arranged protective measures against insulation fault on the building power supply (such as ground fault circuit breaker) may not principally identify possible insulation defects on the output side due to the galvanic insulation through the IMED isolating transformer. This leads to a potential danger for the patients or the operating personnel.
The ELG close this security gap reliably by clearly warning against insulation faults. For this, the leakage resistance is controlled for the undercut of a limit value. If this case occurs, the ELG warns optically as well as acoustically.

 

Specyfikacja techniczna
ELG
1
50 - 60 Hz
230 V
IP40
Wymiary
193 mm
33 mm
56 mm
0,4 kg
Informacje o zamówieniu
9-059-000010
5703451014033

Pliki do pobrania

Brak dostępnych plików do pobrania

Rysunek wymiarowy

No dimension drawings available

Schemat połączeń

Brak schematów połączeń

Produkty z tej kategorii
nazwa Numer artykułu moc Wtórny prądu Waga IP Zapas
ELG 9-059-000010 0,4 kg IP40